東芝CT機(jī)spellman電源維修CT42 CT70北京
SPELLMAN高壓電源廣泛應(yīng)用于東芝CT機(jī)XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代碼及診斷方法大致相同,部分故障代碼排查參考如下
Error contents (note 1) | Cause of problem | Object part | ||
24kW | 48kW/60kW | |||
E13 | Arc- | Cathode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HV | TUBE, INV/HVC |
E25 | Arc+ | Anode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HVA | |
E12 | Over kV- | Cathode side overvoltage. Abnormal main circuit control. | INV/HV, SCB | INV/HVC, SCB |
E27 | Over kV+ | Anode side overvoltage. Abnormal main circuit control. | INV/HVA, SCB | |
E62 | Over mA- | Cathode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HV, SCB | TUBE, INV/HVC,SCB |
E63 | Over mA+ | Anode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HVA,SCB | |
E66 | MAC error | Filament current abnormal. Trouble of the filament heatig circuit. | INV/HV, SCB | INV/HVC, SCB |
E0 | Regulation error (HV abnormal) | There is a difference in binding output and real output | INV/HV, AC/C | INV/HVC, INV/HVA, AC/C |
spellman電源維修CT機(jī)高壓電源維修CT70??PN50X4218北京
醫(yī)療設(shè)備高壓電源維修高壓發(fā)生器維修北京天浦正達(dá)高壓電源專業(yè)維修
Spellman高壓電源XRM/PCM50/70維修射頻電源北京
Spellman高壓電源XRM50P50X3768維修射頻電源北京
spellman高壓電源維修高壓發(fā)生器控制單元維修北京
醫(yī)療設(shè)備高壓電源維修高壓發(fā)生器維修北京天浦正達(dá)高壓電源專業(yè)維修